Hartmann Veroval compact BPW 22 Mode D'emploi page 49

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects
and people.
Table 5
Guidance and manufacturer's declaration - electromagnetic Immunity
The Veroval
®
BPW 22 is intended for use in the electromagnetic environment specified below. The customer or the user of the Veroval
assure that it is used in such an environment.
Test Frequency
(MHz)
Radiated RF
385
IEC61000-4-3
(Test specifications
for ENCLOSURE PORT
450
IMMUNITY to RF wire-
less communications
equipment)
710
745
780
810
870
930
Band a)
Service a)
(MHz)
380 –390
TETRA 400
430 –470
GMRS 460,
FRS 460
704 – 787
LTE Band
13, 17
800 – 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
49
Modulation b)
Modulation b)
(W)
Pulse
1,8
modulation b)
18 Hz
FM c)
2
± 5 kHz deviation
1 kHz sine
Pulse
0,2
modulation b)
217 Hz
Pulse
2
modulation b)
18 Hz
®
BPW 22 should
Distance
IMMUNITY TEST LEVEL
(m)
(V/m)
0,3
27
0,3
28
0,3
9
0,3
28

Publicité

Table des Matières
loading

Table des Matières