Table 2
Guidance and manufacturer's declaration - electromagnetic Immunity
The Veroval
BPW 22 is intended for use in the electromagnetic environment specified below. The customer or the user of the Veroval
®
assure that it is used in such an environment.
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
IEC 60601 Test level
±8 kV contact
±15 kV air
Power supply lines: ±2 kV
input/output lines: ±1 kV
line(s) to line(s): ±1 kV.
line(s) to earth: ±2 kV.
100 kHz repetition frequency
5% U
(>95% dip in U
) for 0.5 cycle
T
T
40% U
(60% dip in U
) for 5 cycle
T
T
70% U
(30% dip in U
) for 25 cycle
T
T
<5% U
(>95% dip in U
) for 5s
T
T
Compliance
Electromagnetic environment - guidance
level
±8 kV contact
Floors should be wood, concrete or ceramic tile. If floors are
±15 kV air
covered with synthetic material, the relative humidity should
be at least 30 %
Not applicable
Mains power quality should be that of a typical commercial
or hospital environment.
Not applicable
Mains power quality should be that of a typical commercial
or hospital environment.
Not applicable
Mains power quality should be that of a typical commercial
or hospital environment.
45
BPW 22 should
®