Hartmann Veroval compact BPW 22 Mode D'emploi page 46

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Immunity Test
Power frequency (50/60Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mians voltage prior to application of the test level.
T
Table 3
Guidance and manufacturer's declaration - electromagnetic Immunity
The Veroval
BPW 22 is intended for use in the electromagnetic environment specified below. The customer or the user of the Veroval
®
assure that it is used in such an environment.
Immunity Test
IEC 60601 Test level
Conduced RF
150KHz to 80MHz:
IEC61000-4-6
3Vrms
6Vrms (in ISM and
amateur radio bands)
80% Am at 1kHz
IEC 60601 Test level
30 A/m
50Hz/60Hz
Compliance level
Not applicable
Compliance
Electromagnetic environment - guidance
level
30 A/m
Power frequency magnetic fields should be at levels
50Hz/60Hz
characteristic of a typical location in a typical commercial
or hospital environment.
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should be used no closer
to any part of the Veroval
BPW 22, including cables, than the recommended
®
separation distance calculated from the equation appropriate for the
frequency of the transmitter. Recommended separation distances:
d = 1.2
;
d = 2
46
BPW 22 should
®

Publicité