18. Electro Magnetic Compliance (EMC)
The Medical Bed is intended for use in the electromagnetic environment specified below. The user of the Medical
bed should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration – electromagnetic immunity
The Medical Bed is intended for use in the electromagnetic environment specified below. The user of the Medical
should assure that it is used in such an environment.
IMMUNITY
IEC 60601
test
test level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
± 2 kV for power
Electrical fast
supply lines
transient/burst
± 1 kV for input/
IEC 61000-4-4
output
lines
± 1 kV line(s) to
Surge
line(s)
IEC 61000-4-5
± 2 kV line(s) to earth
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
Voltage dips, short
40 % UT
interruptions and
(60 % dip in UT)
voltage variations
for 5 cycles
on power supply
70 % UT
input lines
(30 % dip in UT)
IEC 61000-4-11
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Power frequency
(50/60 Hz)
3 A/m
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
Compliance
Electromagnetic environment – guidance
The Medical Bed uses RF energy only for its internal
function. Therefore, its RF emissions are very low and
Group 1
are not likely to cause any interference in nearby
electronic equipment.
Class B
The Medical Bed is suitable for use in all
establishments, including domestic establishments and
Class A
those directly connected to the public low-voltage
power supply network that supplies buildings used
for domestic purposes.
Complies
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 1 kV differential
mode
± 2 kV common mode
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
0,3 A/m
27
Electromagnetic environment –
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
Medical bed Alegio requires
continued operation during power
mains interruptions, it is recommended
that the Medical Bed Alegio be
powered from an uninterruptible
power
supply or a battery.
The power frequency magnetic field
should be at a characteristic level of a
typical commercial or hospital
environment.