2.8.2
Rated data
For rated data please observe section 2.7.2.
2.8.3
Special conditions
In type of protection intrinsically safe apparatus, the analytical probes
(pH ISM sensor simulators) equipment is designated with the
following specific conditions of use.
1. The pH ISM sensor simulators shall be installed in compliance
with the enclosure, mounting, spacing and segregation require-
ments of the ultimate application, including a tool removable
cover.
2. Process temperature no greater than +130 ºC.
3. Potential Electrostatic Charging Hazard – To prevent the risk of
electrostatic sparking, the non-metallic surface should only be
cleaned with a damp cloth.
pH ISM Qualification Kit
52 300 439 E
© 12/2018 Mettler-Toledo GmbH, Greifensee
pH ISM Qualification Kit
Printed in Switzerland