Special Conditions - Mettler Toledo 52 300 411 Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12
2.7.3

Special conditions

– The relationship between the maximum permissible ambient
or media temperature and temperature class, for category 1G
applications, zone 0, is shown in the following table :
Temperature class
T 6
T 5
T 4
T 3
– The capacitance and inductance of the connecting cable has to
be considered.
pH ISM Qualification Kit
52 300 439 E
Max. environment resp.
media temperature
62 °C
74 °C
102 °C
154 °C
© 12/2018 Mettler-Toledo GmbH, Greifensee
pH ISM Qualification Kit
Printed in Switzerland

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

52 300 41252 300 41352 300 41452 300 415

Table des Matières