The speed on the IKA
®
RW 47 D agitator is adjus-
ted by means of the rotary button (C).
You can read the set speed from the marking on
You can select one of the three gear positions
available (0, I or II) using the toggle button (D) on
the front of the appliance. You may only switch
to a different cog wheel gear
The laboratory stirrer IKA
secured to a stable stand with four screws (e.g.
R 474 or R 472 ) and wired according to the wi-
ring plan for proper use. The stirring bowl must
always be well secured for safety reasons.
You must also ensure that the mounting equip-
ment (e.g. fl oor stand, telescope stand) is set up
in such a way that it is not liable to topple and
does not start to move during the stirring proce-
dure. Accessories must be assembled and wired
according to the assembly instructions.
Check whether the voltage given on the type
plate corresponds to the available mains volta-
ge. The socket used must be earthed (fi tted with
earth contact). If these conditions have been met,
the machine is ready to operate when the mains
plug is plugged in. If these conditions are not
met, safe operation is not guaranteed and the
machine could be damaged.
When working with safety switch SI 400, the ap-
pliance only starts up once it has been set to the
designated working height and the switch con-
tact activates the safety switch.
During commissioning of the machine the output
shaft starts to run at the last speed set. Therefore
check the setting of the control knob (speed).

Speed display

Gear speed change

Commissioning
RW 47 D must be
®

Switching on the device

the top of the casing on the label attached to the
motor support. (The values are only valid for low
agitator shaft lows).
level when the appliance is switched off (motor
stopped). Otherwise this may cause damage to
the gear cogs. The arrow on the front plate indi-
cates the direction of the chuck.
If the stirring unit is operated in such a way that
the appliance's EMERGENCY OFF switch (B) (Fig. 2)
is not accessible, you must add an another EMER-
GENCY OFF switch to the work area so that it can
be easily reached.
Points of particular danger (such as crushing, ro-
tating parts, clamping screws for stand legs and
rollers etc.) are marked in Fig. 3 with an exclama-
tion mark. Please pay particular attention to these
danger points during operation!
Also ensure that the speed set is suitable for the
test texture selected. If you are unsure, set the
motor slide to the lowest speed front motor slide
stop) using the side knob and the gear to the
lowest level (Fig. 2).
The appliance will start up when you press the
ON switch on the motor (Fig. 2).
If the unit has not been used for some time, a
knocking noise will be heard when switching
on, which is due to the preload on the friction
lining of the friction wheel drive. This has no de-
trimental effect on the function of the unit, and
the knocking noise will disappear after a short
running-in period.