Accessories; Technical Data - IKA RW 47 D Mode D'emploi

Masquer les pouces Voir aussi pour RW 47 D:
Table des Matières

Publicité

R 472
Floor stand
R 474
Telescopic stand
R 302
Stirring shaft protection
Speed range
50 Hz stage I
50 Hz stage II
60 Hz stage I
60 Hz stage II
Speed adjustment
Speed display
Max. torque stirrer shaft (overload)
Max. torque stirrer shaft
Max. torque stirrer shaft
Motor data
Design voltage
Design frequency
Motor speed
Input power
Output power
Maximum output at stirrer shaft
Permitted on-time
Protection type to DIN EN 60529
Overvoltage category
Protection class
Contamination level
Protection on overload
Perm. ambient temperature
Perm. humidity (rel.)
Operating position
Drive
Maximum stirring quantity (water)
For high-viscosity (HV)
Clamping chuck - clamping range
Hollow shaft internal diameter
Stand attachment
Housing material
Dimension without extension arm
(WxDxH)
Weight
Operation at a terrestrial altitude
Subject to technical changes!

Accessories

SI 400
SI 472
SI 474

Technical data

rpm
57-275
rpm
275 – 1300
rpm
69 - 330
rpm
330 – 1560
Mechanical rotary button on the side
Speed label on the side of the device
Ncm
4642 ( bei 57 rpm )
Ncm
3000 ( bei 100 rpm )
Ncm
285 ( bei 1000 rpm )
VAC
3 x 400 Y
Hz
50
rpm
2880
W
554
W
370
W
300
%
100
IP 44
II
I (Protective earth)
2
overcurrent and undervoltage switch
on the three-phase motor
°C
+ 5 bis + 40
%
80
On stand, clamping chuck pointing down
Rib cooled three-phase motor with friction
wheel drive and subsequent 2-stage toothed
gear train
l
200
mPas
10000-100000
mm
3-16
mm
13 (upto short before the internal housing
upper edge)
Flange on housing lower part
Al, hermoplastic
mm
145x445x340
kg
15
m
max. 2000
Safety limilt switch
Support holder
Support holder
3 x 230 ∆
3 x 400 Y 3 x 230 ∆
50
60
2850
3425
548
489
370
370
60
3425
490
370
17

Publicité