• Gradually increase the shaking rate..
• Reduce the shaking rate if
- the medium splashes out of the vessel because the speed is
too high
- the appliance is not running smoothly
• If you notice that the device is not running smoothly, the
shaking rate must always be reduced until no more uneveness
occurs in the operation.
• Because of improper loading and the position of the center of
gravity, dynamic forces may arise during the agitation process
that cause the shaker to move about on the table.
For restrictions of load capacity and material weight during
high shaking frequencies, please see the description in the
„Attachments" section.
• Pay attention to the vessels on the shaking table when setting
the shaking rate. This will prevent any of the medium to be
shaken from spurting out of the sample vessels.
• Should vessels break during operation or the media be spilt,
interrupt the shaking process immediately, remove any vessel
residue and clean the appliance.
• Never touch moving parts.
Caution! The shaker still runs after the
DANGER
cover has been opened.
Wait for it to stop running.
(Risk of being crushed, shocked or cut, see figure illustrating
danger points).
Correct use
• U se
- for mixing and / or heating liquids
The device is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply network that also supplies residential areas.
The safety of the user cannot be guaranteed if the appliance is operated with accessories that are not supplied or recommended by
the manufacturer or if the appliance is opreated improperly contrary to the manufacturer's specifications or if the appliance or the
printed circuit board are modified parties.
Unpacking
• Contents of package
KS 3000 i control
- Shaking device
- Mains cable
- Operating instructions
- 4 clamping screws short
- Temperature sensor PT 1000.80
- USB-Cable
- RS232-Cable
- safety declaration
Accessories
• NOTE ! Covering or parts that are capable of being removed
from the unit without accessory equipment have to be re-
tached to the unit for safe operation in order to prevent, for
example, the ingress of fluids, foreign matter, etc..
• All accessories and vessels in place for the shaking process
must be firmly secured.
• Accessories may only be assembled once the plug has been dis
connected. The safety of the user cannot be guaranteed if the
appliance is operated with accessories that are not supplied or
recommended by the manufacturer or if the appliance is opera-
ted improperly, contrary to the manufacturer's specifications.
• Shaking vessels which are not properly secured could get
damaged or be projected out, thus causing injury. It is essenti-
al to regularly check that the vessels to be shaken and the
attachments are firmly secured, especially before using the
appliance again.
• When using an external temperature sensor, it must always be
in the medium. Immerse the external temperature sensor at
least 20mm into the medium.
To the protection of the equipment
• When servicing, the wiring selected by IKA must be set up
again!
• Avoid allowing objects to push or strike the agitation table.
• Keep a minimum distance of 100mm from the appliance on all
sides.
KS 3000 ic control
- Shaking device
- Mains cable
- Operating instructions
- USB-Cable
- 4 clamping screws shor
- Temperature sensor PT 1000.80
- 2 hose connecting pieces
- 2 hose connecting clamp
- 2 unlocking handle
- RS232-Cable
- safety declaration
• Range of use
- Laboratories
- Schools
- Pharmacies
- Universitys
• Unpacking
- Please unpack the device carefully
- In the case of any damage a
detailed report must be set
immediately (post, rail or
forwarder)
15