Factory Settings; Correct Use - IKA TRON ETS-D5 Mode D'emploi

Masquer les pouces Voir aussi pour TRON ETS-D5:
Table des Matières

Publicité

The safety advice given above represents the current status of the
known risks. Nonetheless, the user must check equipment for
correct and safe operation before operating any system compri-
sing the ETS-D5 and a magnetic agitator/heating system.

Correct use

The IKATRON ETS-D5 is an instrument for
the precise control of temperature. The sensor
on this instrument must be inserted at least
20mm into the medium to be tempered.
The instrument can be connected to any
magnetic agitator or heating plate that has
a contact thermometer connection, provided
this conforms with the requirements listed
in the technical data
(see section Technical Data).
Unpacking
Please unpack the equipment carefully and inspect for damage. It
is important that any transit damage should be identified at the
time of unpacking. If necessary the equipment should be returned
immediately for inspection (by post, rail or freight forwarder).
As supplied, the package contains: one IKATRON ETS-D5 and one
set of operating instructions
.
10

Factory settings

The ETS-D5 is supplied with the following factory settings:
Operating mode
A
Set-point temperature
-10 °C
HI-TEMP
450 °C
Restoring parameter to favtory settings (RESET)
The following steps must be followed to restore factory settings:
• Switch off the equipment
• Press and hold down the membrane keys "L" and "M" and
switch on the equipment
• The factory settings will now be restored (see above)
Installation
Plug the instrument into the contact thermometer jack on the
magnetic agitator (e.g. IKA RCT basic, IKA RET basic, RH basic,
RH digital KT/C, Heidolph MR3001K, Heidolph MR3002, VWR
VMS-D, VWR VMS-A, H+P Variomag, etc.).
Please observe all safety instructions in this respect.
Procedure for connecting a Corning Magnetic Agitator:
Turn the rotary "heat" dial on the magnetic agitator to the " " O O " " (off)
position. Next, connect the ETS-D5 to the magnetic agitator using
the AD-C1 adapter provided.
Please observe all safety instructions in this respect.
The device will be activated when the magnetic agitator is swit-
ched on. If the supply voltage at the contact thermometer connec-
tor is not between 8V and 16V DC then it will be necessary to
power the device using the H52 power supply set (H50 double
connector adapter and H51 plug-in power supply unit). The power
supply and/or double connector adapter will provide the necessary
supply voltage for the device. An automatic self-test will be per-
formed when the device is activated. All segments on the LCD
display will be visible during this test (Figure 1).
ETSD5 0306