Beurer EM 29 Mode D'emploi page 119

Masquer les pouces Voir aussi pour EM 29:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 30
Table 3
Guidance and manufacturer's declaration – electromagnetic immunity
The EM 29 is intended for use in the electromagnetic environment specified below. The customer or the user of the EM 29 should
assure that it is used in such an environment.
IMMUNITY test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power sup-
ply lines
± 1 kV line to line
± 2 kV line to earth
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 s
3 A/m
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power sup-
ply lines
± 1 kV line to line
± 2 kV line to earth
<5 % UT (>95 % dip in UT)
for 0.5 cycle
40 % UT (60 % dip in UT)
for 5 cycles
70 % UT (30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT)
for 5 s
Not applicable
119
Electromagnetic environment – guidance
Floors should be wood, concrete or cera-
mic tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Not applicable
Not applicable
Not applicable
Not applicable

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Tens em 29

Table des Matières