Guidance and manufacturer's declaration – ELECTROMAGNETIC IMMUNITY
FOR DEVICES AND SYSTEMS THAT ARE NOT OF VIABLE FUNCTION
The MAG2000 Family is expected to operate in the electromagnetic environment
below specified. The user or operator of the MAG2000 family must ensure that it
is used in this environment.
Immunity test
Portable and mobile RF communications equipment should not be used near any
part of the MAG2000 family, including cables, except where recommended
separation distances are observed, calculated from the equation applicable to
frequency of the transmitter.
Conducted RF
IEC 61000-4-6
Irradiated RF
IEC 61000-4-3
Irradiated RF for
radio
communication
devices
IEC 61000-4-3
where (P) is the maximum output power rating of the transmitter in Watts (W) according to the
transmitter manufacturer and (d) is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey
, should be less than the compliance level in each frequency range
a
Interference may occur in the vicinity of equipment marked with the following
symbol:
Note:
(1)
At 80 MHz and 800 MHz the separation distance for the higher frequency range
applies.
IACER Srl
Test level
Compliance
IEC 60601
3V
eff
from 150 kHz
3V
eff
to 80 MHz
([V
6V
eff
6V
eff
from 150 kHz
([V
to 80 MHz for
ISM band
10V/m
10V/m
from 80 MHz
[E
1
to 2,7 GHz
3 V/m
3V/m
from 80 MHz
[E
1
to 6 GHz
Electromagnetic
level
environment – guidance
Recommended separation distance
] V)
3
5 ,
1
=
d
V
1
for ISM band
] V)
1
12
=
d
E
1
from 80 MHz to 800 MHz
] V/m
7
=
d
E
1
from 800 MHz to 2,7 GHz
6
=
d
E
] V/m
1
from 80 MHz to 6 GHz
35
12
=
=
P
d
P
V
1
P
P
P
.
b
MNPG55-07