Preventative Maintenance; Checklist - Stryker Stair-PRO 6250 Manuel D'utilisation Et D'entretien

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 45
Operation
Cleaning and Disinfecting
Inspection
Stair-TREAD™ system belt
reconditioning (6252)
Note: Keep up to date maintenance records using the Maintenance Record form on
Parts, Service or Technical Assistance
Contact Stryker Customer Service at 1-800-327-0770 or
Stryker Medical
3800 E. Centre Ave. Portage, MI 49002
ATTN: Customer Service

CHECKLIST

_____ All fasteners secure (reference all assembly drawings)
_____ All welds intact, not cracked or broken
_____ No bent or broken tubing or sheet metal
_____ No debris in wheels
_____ All wheels secure and rolling properly
_____ Front casters secure, rolling and swiveling properly (if equipped)
_____ Wheel locks hold wheels securely when on and clear the wheels when off
_____ Chair unfolds and locks properly
_____ No rips or cracks in seat or back rest
_____ Restraint straps intact and working properly
_____ Foot end carrying handles extend and lock properly
_____ Head end carrying handles fold and unfold
_____ Upper control handle extends and locks in all positions (if equipped)
_____ Stair-TREAD™ system mechanism unfolds and locks properly (if equipped)
_____ Stair-TREAD™ system belt rolls properly (if equipped)
_____ Stair-TREAD™ system belt inner cords not showing - replace if necessary (if equipped)
_____ Stair-TREAD™ system performs as desired - recondition belts if necessary (if equipped) (see
_____ No lubricants present on the Stair-TREAD™ system belts or the track frame surfaces
_____ Upper release handle cable not worn or frayed - replace, if necessary (Model 6252)
_____ Optional accessories intact and operating properly
Serial Number:
Completed by: _______________________________________
www.stryker.com

Preventative Maintenance

Schedule
Each Use.
For 1-25 calls per month, inspect chair every 6 months
For 26-200 calls per month, inspect chair every 3 months
For 201+ calls per month, inspect chair monthly
After usage on approximately 500 flights of stairs or if the
Stair-TREAD™ system performance is sluggish
6250-001-161 REV D
Procedure
See
page 1-30
See below for
checklist
See
page 1-39
page 1-32.
page
1-39)
Date: _________________
Return To Table of Contents
English
1-31

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Stair-pro 6251Stair-pro 6252

Table des Matières