4
Tenupol-3
Instruction Manual
1. Installation
1.1. Packing List
1
Control unit with cable for connection to main power supply
1
Polishing unit with polishing cell, cover for polishing cell, built-in
IR-transmitter/receiver, cooling coil and pump system as well as cable
and plug connector for control unit
1
Thermometer pocket
1
Thermometer +20/-50°C
1
Specimen holder for 03 mm specimens
1
Jet holders, inside diameter 1 mm, set with 2 pcs.
1
PVC-reservoir, insulated
1
PVC-reservoir, uninsulated
2
Ø8 mm tube for cooling coil, 2x1 m.
1
Jet holder with ascending tube
1
Return tube
1.2. Placing
Tenupol-3 is a stationary instrument for setting up in a laboratory. This
implies that the instrument must be placed on a plane and horizontal table.
The table must be resistant to chemicals. It is recommended to place the
polishing unit on a piece of impregnated paper or some other easily
replaceable protective foil.
Please note that the electrolytes to be used may result in demands for local
exhaust or installation in fume cupboard.
The control unit can either be placed next to the polishing unit or on a shelf
above the polishing unit. If placed on a table next to the polishing unit the
operating convenience can be increased by placing the control unit with the
built-in legs in an inclined position.