Icons missing from the
display
Calibration procedure
failed or can not be
completed
Icon has been
de-selected
Turbine may be
faulty
Turbine not fitted
tightly to
calibration
syringe
Calibration
syringe does not
have an
inspiratory seal
or seal is leaking.
Shaft of the
syringe being
pushed down
37
Hold stylus on to the blue
area of the display, a list
will appear, ensure
required icon is selected.
Repeat calibration
procedure, if problem
still persists, replace
turbine or return unit for
servicing
Ensure the syringe is
fitted to the turbine using
the syringe adapter
supplied, part #
(supplied with the
syringe)
Ensure you are using
manufacturers
recommended syringe.
The syringe should be
emptied and filled with
one smooth stroke, avoid
pushing down on the
shaft or banging at the
end of each manoeuver.
861427.