Normal pressure lung demand valve with-
out medium pressure coupling
Preparing the test device
Necessary accessories:
— Test piece PE (R 51 591)
— Test piece P (R 50 028)
— ESA adaptor (AG 02 406)
— Adaptor for lung demand valve of the PSS 90/100
(33 37 650)
— Adaptor for lung demand valve of the PSS 7000 (on
request)
Close the compressed air cylinder of the compressed
air breathing apparatus.
Remove the lung demand valve from the pressure
reducer.
Fit the adaptor on the connection of the lung demand
valve.
Fit the adaptor into the pressure reducer.
Detach the test head.
Connect the lung demand
valve with the corresponding
test piece/adaptor to the round
thread connection (1).
Open the compressed air sup-
ply.
Activate the measuring point "Round thread connec-
tion" (2).
Check for leaks
1)
Set the test time
(1).
Disconnect the hose with a
blunt tool.
Generate the test pressure
Briefly pull the joystick (2) for
fine adjustment forwards sev-
eral times.
The pressure is displayed on the low pressure
manometer (3).
Start the timer (1).
Read the pressure at the end of the test time and com-
pare the pressure difference with the target value
Testing the opening pressure
Slowly generate a negative
pressure (1).
Observe the manometer (2).
Read the opening pressure
and compare with the target
1)
value
.
1) Refer to the Instructions for Use of the lung demand valve.
1
2
3
1
1)
2
:
1)
.
2
1
Testing the metering valve
1)
Set the test time
(1).
Generate the test pressure
The pressure is displayed on
the low pressure manometer
(2)
Start the timer (1).
Read the pressure at the end of the test time and com-
pare the pressure difference with the target value
Ending the test
Depressurize the line system.
Remove the test assembly.
1)
.
2
1
1)
.
27