Maintenance; Recommended Maintenance Intervals; Replacing The Battery Of The Timer; Disposal - Dräger Testor 2100 Notice D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 34

Maintenance

Recommended maintenance intervals

Work to be carried out
Perform self test
Clean test device
Functional check and leak test
Check accuracy of manometer
Calibrate high, medium and low
pressure sensors
Fully overhaul high pressure com-
ponents
1)
Carried out by Dräger
Recheck for functionality and leakages after servicing
and/or replacing components.

Replacing the battery of the timer

 Disconnect the timer from the
test device (attached with a
Velcro strip).
 Open the battery compart-
ment:
Use the suitable screwdriver or
coin, turn in a counterclock-
wise direction and remove the cover.
 Remove the old battery and insert the new battery.
Ensure the correct polarity, "+" must face upwards.
 Close the battery compartment:
Attach the cover and close it using the suitable screw-
driver or coin.
 Attach the timer with the Velcro strip to the test device.

Disposal

EU-wide regulations for the disposal of
electric and electronic appliances
which have been defined in the EU
Directive 2002/96/EC and in national
laws are effective from August 2005
and apply to this device. Common
household appliances can be dis-
posed of using special collecting and
recycling facilities.
X
However, as this device has not been registered for
household usage, it must not be disposed of through
X
these means. The device can be returned to your national
Dräger Sales Organization for disposal. Please do not
X
hesitate to contact the above if you have any further
questions on this issue.
1)
X
1)
X

Troubleshooting

1)
Fault
X
Test head looses
air
Positive pres-
sure in mask
Test pressure not
reached
Timer not work-
ing
Cause
Remedy
Test head leaky
Retighten con-
nector clamp
Check hose con-
nections
Mouth nozzle on
Send for repair to
test head leaky
Dräger
Supply pressure
Check supply
too low
pressure; fill cyl-
inder as required
Adaptor incorrect
Check
or leaky
Battery empty
Replace battery
31