TROUBLE SHOOTING

Error Message

Fault
bAtt
The main battery module is low.
_ _ _ _ _
An error has occurred during the measurement.
e.g. contact has been lost at one of the probes.
ERR 114
Checksum Failure in EEPROM.
Calibration constants have been lost.
The DLRO will continue to work but the accuracy will now
be typically ± 2%. At switch on the display will show the
software version with dashes each side e.g. - 1.0 -
ERR 115
Checksum Failure in battery backed RAM.
This can occur if the main battery module and the
backup battery become completely exhausted.
It is possible to reset your DLRO 10X to the original factory settings. This will clear all stored results, set the clock to a default value and format, and will set
all other settings to their default values. This is effected by holding down the '3' button on the alphanumeric keypad while switching on your instrument.
Action required.
Recharge the main battery or replace with a charged one
Rectify the error and repeat the measurement.
Return for recalibration.
Recharge the main battery or replace it with a charged
one. Switch the instrument on.
Calibration constants will be retrieved from EEPROM.