Delete Data; Storage; Range Menu; Entering Notes In The Memo Screen - Megger DLRO-10 Guide De L'utilisateur

Masquer les pouces Voir aussi pour DLRO-10:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

GENERAL OPERATION - DLRO 10X
The row below DD MM YY HH MM contains respectively the date, the
month, the two-digit year, the hour of the day in 24-hour notation and the
minute. These must be entered in this sequence regardless of the date
format you wish to use.
TEST
OPTIONS
NORM
DD MM YY HH MM
09/10/00 15.08
DD MM YY
09/10/00 12.02
The lower DD/MM/YY shows the current date format. Pressing the cursor
control Up arrow will cycle through the available options DD/MM/YY, MM/
DD/YY or YY/MM/DD.
The bottom line shows the current setting of date and time. This is
updated when you press Enter to exit the Set Clock function. However,
tests already stored prior to changing the date format will retain the old
format.

Delete Data

Select Delete Data if you wish to clear DLRO 10X's memory of stored data.
In case you have selected this Option by accident you will be asked to
confirm that you wish to delete the data; the default is NO. Change this to
YES and press Enter if you wish to delete all data.
Please note - ALL stored data will be deleted.

Storage

The Storage option sets the default for data storage. It may be set to
always store data or never to store data. Highlight the desired option at
RANGE
AUTO
the bottom of the screen and press Enter.
With the exception of tests carried out in Continuous mode, at the end of
each test you have the option to change the default by selecting STORE or
NO STORE at the bottom of the results screen.
At the end of a test, if memory is full this STORE / NO STORE message will
change to MEMORY FULL and no more tests will be stored.
Data is stored indefinitely if a charged battery pack is fitted. If the main
battery becomes exhausted or is removed, data is maintained for a period
of 4 days by an internal backup battery. When a charged battery pack is
refitted, the backup battery will recharge to full capacity within 1 week.

Range Menu

DLRO 10X uses a test current between 100 µA and 10 A to measure the
resistance of the item under test. If set to AUTO, DLRO 10X will select the
current according to the resistance it detects. (see Specifications, Ranges)
However, in some cases it may be desirable set a maximum test current. If
this is so, move the cursor to the desired current and press Enter.
Six test currents are available; 10 A, 1 A, 100 mA, 10 mA, 1 mA and 100 µA.
Selecting the 10 A maximum will have the same effect as selecting AUTO.
The Main Menu screen will show the active current range beneath the
RANGE heading.

ENTERING NOTES IN THE MEMO SCREEN

At the end of each test, (with the exception of a test in Continuous mode),
you can select the STORE or NO STORE option to store the test results or
not. Select the desired option using the cursor control. This overrides the
default setting under Options, Storage for one test only.
9

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

Dlro-10xDlro 10

Table des Matières