Cleaning and disinfecting the drinking valve
●
Prepare a cleaning solution: Completely dissolve 1 cleaning tablet
R 57 991 in 33.8 US fl.oz. (1 liter) of drinking water and fill it into
the drinking reservoir.
●
Remove the mouthpiece and replace with the drain hose.
●
Clean the mouthpiece in the cleaning solution. Rinse with running
drinking water and dry with oil-free compressed air (10 - 20 psi/
0.6 - 1.2 bar).
●
Connect the drinking reservoir to the outside of the drinking valve
with a hose.
●
Squeeze the cleaning solution from the reservoir through the
drinking valve until a small amount of the solution emerges from
the drain hose.
●
Close off the drain hose with a hose clip.
●
Allow the cleaning solution to act for 3 minutes.
●
Open the hose clip and squeeze roughly half of the cleaning
solution out of the reservoir and through the drinking valve.
●
Close the hose clip again and let the cleaning solution act for
3 minutes again.
●
Open the hose clip and squeeze the remaining cleaning solution
through the drinking valve.
●
Fill the drinking reservoir with clear drinking water.
●
Squeeze the water out of the reservoir and through the drinking
valve.
●
Disconnect the drinking reservoir from the hose.
●
Using a compressed air gun, blow oil-free compressed air (10 -
20 psi/0.6 - 1.2 bar) into the hose until the drinking valve is dry.
●
Dismantle the hose and the drain hose.
●
Fit the mouthpiece onto the connection inside the facepiece and
make sure that it is properly fixed.
138
Leak tests
Carry out the leak test with a suitable test device (e. g. of the Testor
or Quaestor series)
(1) Checking leak tightness
●
Moisten the face seal with water.
●
Mount the Dräger FPS 7000 RP on the test head.
●
Make sure that the face seal fits tightly onto all parts of the test
head. Inflate the test head if necessary.
●
Adjust the connection piece with adapter tightly and produce a
negative pressure of 10 cm H
If the pressure drop is less than 1 cm H
Dräger FPS 7000 RP is considered leak-tight. If the Dräger
FPS 7000 RP leaks, continue with step (2).
(2) Leak test under water
●
If fitted, remove the communication system.
●
Immerse the test head including the Dräger FPS 7000 RP in the
water.
●
Produce a positive pressure of approx. 10 cm H
●
Slowly turn the test head under the water surface. Emerging
bubbles indicate leaks.
●
Take the test head with the Dräger FPS 7000 RP out of the water,
repair the Dräger FPS 7000 RP or replace leaking components.
●
Repeat the leak test (1).
After testing
●
Press the button and pull the adapter from the connection piece.
●
Remove the Dräger FPS 7000 RP from the test head and dry if
necessary.
Use Instructions for NIOSH-approved full facepieces
O (10 mbar).
2
O (1 mbar) per minute, the
2
O (10 mbar).
2