10 - ELECTROMAGNETIC COMPATIBILITY

Warning: The charger power cord must be kept apart from those of any nearby devices.
M
L.E.D. requires special precautions to be taken with regard to electromagnetic compatibility. It must be installed
INI
and prepared for use as described in chapter 3.
Certain types of mobile telecommunication devices such as mobile telephones are likely to interfere with the M
L.E.D.. The recommended separation distances in this paragraph must therefore be complied with.
The M
L.E.D. must not be used near or on top of another device.
INI
If this cannot be avoided, its operation under the conditions of use must be checked beforehand.
The use of accessories other than those specified or sold by SATELEC as replacement parts may have the consequence
of increasing the emissions or decreasing the immunity of the M

10.1 - Electromagnetic emissions

The M
L.E.D. is intended for use in the electromagnetic environment specified in the table below. The user and/or
INI
installer must ensure that the M
Emission test
Compliance
RF emission - CISPR
11.
10.2 - Magnetic and electromagnetic immunity
The M
L.E.D. O
is intended for use in the electromagnetic environment specified in the table below.
INI
RTHO
The user and/or installer must ensure that the M
Immunity test
Magnetic field at 50Hz.
3A/m
IEC61000-4-8
Electrostatic discharge
± 6KV contact
(ESD)
± 8KV air
IEC 61000-4-2.
Electrical fast transients
± 2KV for power supply
IEC 61000-4-4.
lines
Surges
± 1KV differential mode
IEC 61000-4-5.
± 2KV common mode
<5% UT
0.5 cycles.
Voltage dips, short
40% U
interruptions and voltage
cycles
variations
70% U
IEC 61000-4-11.
25 cycles
<5% U
250 cycles
L.E.D. is used in such an environment.
INI
M
L.E.D. uses RF energy for internal operation.
INI
Therefore, its radiofrequency emissions are very low and are not likely to cause any
interference in nearby equipment.
Group 1
Class B
M
L.E.D. uses RF energy for internal operation.
INI
Therefore, its radiofrequency emissions are very low and are not likely to cause any
interference in nearby equipment.
IEC 60601 test level
3A/m
± 6KV contact
± 8KV air
± 2KV for power supply
lines
± 1KV differential mode
± 2KV common mode
(>95% dip in U
) for
<5% U
T
T
0.5 cycles.
(60% dip in U
) for 5
40% U
T
T
cycles
(30% dip in U
) for
70% U
T
T
cycles
(>95% dip in U
) for
<5% U
T
T
250 cycles
L.E.D..
INI
Electromagnetic environment - guidance
L.E.D. O
is used in such an electromagnetic environment.
INI
RTHO
Compliance level
The intensity of the magnetic field should be equivalent to that of
a typical commercial or hospital environment (hospital, clinic).
Floors must be wood, concrete, cement or tiled.
If floors are covered with synthetic material (carpet, etc.), the
relative humidity must be at least 30%.
Power quality should be that of a typical commercial or hospital
environment.
Power quality should be that of a typical commercial or hospital
environment.
(>95% dip in U
) for
T
T
Power quality should be that of a typical commercial or hospital
(60% dip in U
) for 5
T
T
environment.
If the use of the M
(30% dip in U
) for 25
T
T
power cut, it is recommended that the product be powered from
a separate power supply (UPS, etc.).
(>95% dip in U
) for
T
T
9
Electromagnetic environment - guidance
L.E.D. requires continued operation during a
INI
INI