Analogue Outputs - Camille Bauer EURAX DME 440 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Analogue outputs

For the outputs A, B, C and D:
Output
Impressed
variable Y
DC current
Full scale Y2
see "Ordering
information"
Limits of output
signal for input
overload
and/or
R = 0
1.25 · Y2
R → ∞
30 V
Rated useful
7.5 V
0 ≤
range of output
Y2
load
AC component
≤ 0.005 · Y2
of output signal
(peak-to-peak)
The outputs A, B, C and D may be either short or open-
circuited. They are electrically insulated from each other and
from all other circuits (floating).
System response
Duration of the
measurement cycle:
Approx. 0.5 to 1.2 s at 50 Hz,
depending on measured variable
and programming
Response time:
1...2 times the measurement cycle
Accuracy class:
(the reference value is the full-scale
value Y2)
Meas. variable
Condition
System:
0.5 ≤ X2/Sr ≤ 1.5
Active, reactive and
0.3 ≤ X2/Sr < 0.5
apparent power
Phase:
0.167 ≤ X2/Sr ≤ 0.5 0.25 c
Active, reactive and
0.1 ≤ X2/Sr < 0.167 0.5 c
apparent power
0.5Sr ≤ S ≤ 1.5 Sr,
(X2 - X0) = 2
0.5Sr ≤ S ≤ 1.5 Sr,
1 ≤ (X2 - X0) < 2
0.5Sr ≤ S ≤ 1.5 Sr,
Power factor,
0.5 ≤ (X2 - X0) < 1
active power factor
0.1Sr ≤ S < 0.5Sr,
and reactive
(X2 - X0) = 2
power factor
0.1Sr ≤ S < 0.5Sr,
1 ≤ (X2 - X0) < 2
0.1Sr ≤ S < 0.5Sr,
0.5 ≤ (X2 - X0) < 1
0.1 Ur ≤ U ≤ 1.2 Ur 0.2 c
AC voltage
AC current/
0.1 Ir ≤ I ≤ 1.5 Ir
current averages
0.1 Ur ≤ U ≤ 1.2 Ur 0.15 + 0.03 c
System frequency
resp.
0.1 Ir ≤ I ≤ 1.5 Ir
Energy meter
acc. to IEC 1036
0.1 Ir ≤ I ≤ 1.5 Ir
* Basic accuracy 0,5 c for applications with phase shift
38
Impressed
DC voltage
see "Ordering
information"
40 mA
1.25 Y2
Y2
Y2
15 V
≤ ∞
2 mA
1 mA
Y2
≤ 0.005 · Y2
Accur. class*
0.25 c
0.5 c
0.25 c
0.5 c
1.0 c
0.5 c
1.0 c
2.0 c
0.2 c
(f
= 50...60 Hz)
N
0.15 + 0.1 c
(f
= 16 2/3 Hz)
N
1.0
Factor c (the highest value applies):
Linear characteristic:
Bent characteristic:
X0 ≤ X ≤ X1
X1 < X ≤ X2
Y
X0/Y0
X2/Y2
X
Limit of the output
range
Fig. 7. Examples of settings
with linear characteristic.
Influencing quantities and permissible variations
Acc. to EN 60 688
Safety
Protection class:
II
Overvoltage category:
III
Insulation test:
Input voltage:
Input current:
Output:
Power supply:
Power supply
Voltage:
Acc. to type label
≤ 9 W resp. ≤ 10 VA
Consumption:
Programming connector on transducer
Interface:
RS 232 C
DSUB socket:
9-pin
5
9
GND
The interface is electrically insulated
DTR
CTS
TXD
from all other circuits.
RTS
RXD
DSR
6 1
Y0
1 -
Y2
c =
or c = 1
X0
1 -
X2
Y1 - Y0
X2
or c = 1
c =
·
X1 - X0
Y2
Y1
1 -
Y2
c =
or c = 1
X1
1 -
X2
Y
X0/Y0
X1/Y1
X2/Y2
Limit of the output
range
Fig. 8. Examples of settings
with bent characteristics.
AC 400 V
AC 400 V
DC 40 V
AC 400 V
DC 230 V
X

Publicité

Table des Matières
loading

Table des Matières