7 Conditioning
38
Operating Instructions Midwest Tradition
Please observe the manufacturer's instructions for using instrument
disinfectants.
7.2
Automated cleaning
The cleaning equipment used must be approved by its manufacturer for
the cleaning of dental instruments.
For further details, refer to the operating instructions supplied with the
unit.
✔ The high-speed handpiece is conditioned with a cleaning device.
1. Check whether the high-speed handpiece is clean after conditioning
under good lighting (min. 500 lux) and color rendering index (min. 80
Ra).
2. If they are still dirty, repeat the process.
For further processing, the high-speed handpiece should be dry
and free of residue.
®
®
TB/Pro TBF / Midwest Tradition
TL/Pro TLF