1 • DESCRIPTION
The KINETEC Performa is a Continuous Passive Motion exerciser for EXTENSION-FLEXION mobilization of the HIP
and KNEE joints.
A - Lower limb support • B-Thigh support • C-Foot plate • D - Hand control • E - Power cord
2 • ELECTRICAL CONNECTION: SAFETY FIRST.
The KINETEC Performa is a Type B Class I machine.
Plug the power cord (1) on the KINETEC Performa unit into a receptacle (Voltage between 100 to 240 Vac 50/60 Hz).
IMPORTANT:
Check to be sure that the plug fits the socket being used.
The socket must be grounded.
Use only the power cord delivered with the unit.
Ensure that the power cords are not entangled in the machine during operation.
Fuse change:
Warning: Replace the fuse with one of the same type and value.
3 • SAFETY
The healthcare professional determines the CPM protocol to be used (set up, duration and frequency of each session).
The patient must understand the Start/Stop/Reverse function of the hand control.
This hand control is water proof to the UL 544 standard and should only be repaired by Kinetec.
The KINETEC Performa unit is CE marked (Directive 89/336/CEE)
DANGER: Explosion risk if used with flammable anesthetics.
4 • START UP
Turn on the power switch (2). The hand control will read .
Locking the hand control settings: switch (3)
UNLOCKED POSITION
adjustments are possible.
Double locking: simultaneously press keys PAUSE EXT and PAUSE FLEX for blocked the hand control, the display reads
"LOCK", for unblocked the hand control, simultaneously press the same keys, the display reads "UNLOCK".
We recommend that the switch be kept in the locked position when the hand control is given to the patient.
Display
4 - 3 characters indicate the EXTENSION limit.
5 - 3 characters indicate the FLEXION limit.
6 - 8 characters indicate messages. (RUN, STOP, EXT, FLEX, SPEED, FORCE, PAUSE, etc.).
7 - 3 characters indicate the real angle of the KINETEC Performa.
Key functions
8 - Reading or selecting EXTENSION limit from - 3° to 125°.
9 - Reading or selecting FLEXION limit from 130° to 2°.
10 - Reading or selecting TIME COUNTER from 24H00MIN to 00H00MIN
At the end of the session, the display reads
Depress by + or - to select a new session time.
11 - These keys enable you:
- while selecting, to reduce or increase the operating values (blink value),
- while moving, to exceed the EXTENSION or FLEXION limits (BY-PASS MODE).
- while stopped, to manually move to EXTENSION or FLEXION (MANUAL MODE).
To activate BY-PASS MODE or MANUAL MODE, press the key for at least 3 seconds.
Press the EXTENSION or FLEXION key to memorize the true angle reached, if necessary.
12 - This key enables you to start movement.
12' - This key enables you stop movement, press START again to reverse movement.
IMPORTANT: for maximum safety, the hand control should always be given to the patient.
13 - Reading or selecting SPEED from 7 to 1.
14 - Reading or selecting SYNCHRO-STIMULATOR (please request the instruction sheet from your KINETEC distributor).
15 - Plug for synchro stimulator
(use only approved stimulators to EN 601.2.10 / 1987 standard and/or with EC marked).
16 - Reading or selecting movement inversion LOAD from: _ _ _ _ _ _ = maxi load to _ = mini load.
17 - Reading or selecting PAUSE in EXTENSION or FLEXION limit from 900 seconds (15 minutes) to 0 second.
- RESET TIME function
The machine is at rest. Simultaneously press keys EXTENSION and FLEXION
This function enables you to read the operation time since the lest reset.
Press the key START twice to reset the counter.
5 • SPECIFICATIONS
Product
Weight: 15 Kg (33,04 lbs)
Dimensions: 109cm x 33cm x 35cm
(42.91 x 13 x 13.78 inches)
Range of motion: from -3° to 130°
Speed: from 35 to 180°/min for long femur
from 40 to 215°/min for medium femur
from 65 to 385°/min for small femur
Patient sizing: from 1.12m to 2.06m
(44.1 to 81.1 inches)
USER MANUAL
KINETEC PERFORMA
LOCK
LOCKED POSITION
the OPERATIONAL settings can be read and the Start/Stop/Reverse
function operated.
TIMER 00H00MIN
Electrical
Power supply:
from 100 to 240 V~
Frequency: 50/60Hz
Electric power: 50VA
Current consumption:
150 mA (under 230 V~)
300 mA (under 120 V~)
Device of type B class I
- on main socket, 2 fuses T 500 mA 250V (6.3x32)
- on power supply PCB, 1 fuse F 2 A 250V (5x20)
- on PCB, 1 fuse T 2 A
and the unit stops at an average angle use.
Environmental
- Storage / shipping conditions:
Ambiant temperature: -40°C to +70°C.
Relative humidity up to 90%.
- Operating conditions:
Ambiant temperature: +10°C to +40°C.
Relative humidity up to 80%.
(-40°F to +160°F)
(+50°F to +105°F)