Troubleshooting
24
The data can be deleted from the memory selected by pressing and
holding the MEMORY button
5
the display
shows "
been deleted.
The unit switches off automatically after about 1 minute, if no button is
pressed during this period. It can also be switched off by pressing the
4
START button
. The time and the personal memory selected are
constantly shown in the display
Error during measurement
Lo Pulse is lower than 40 beats/minute
Hi
Pulse / Blood pressure above measurement range
Determining pulse
Unit has detected an arrhythmia
Batteries low. Please replace batteries.
Error messages
Possible causes and solutions
Err 1
Measurement of the systolic blood pressure was completed.
Either the connection with the hose has been interrupted or no pulse
could be determined.
Check the connections between the cuff and the main unit.
Err 2
Non-physiological pressure has been detected.
Cause: the arm was moved during the measurement.
Repeat the measurement, keeping your arm still.
Err 3
If the cuff takes too long to inflate, it is possible that the cuff is incorrectly
positioned or that the connection with the hose is not airtight.
Check the connections and repeat the measurement.
If the same error occurs repeatedly with further readings, please contact
MEDISANA.