Miscellaneous - Medisana MTP Mode D'emploi

Masquer les pouces Voir aussi pour MTP:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 31
6.1 Guide to display symbols
Err
Error during measurement
Lo
Pulse is lower than 40 beats/minute
Determining pulse
Batteries low. Please replace batteries.
6.2 Troubleshooting
If the display shows one of the following error messages, this means that the unit
was unable to accurately determine blood pressure.
Error messages
Possible causes and solutions
ERR 1 Measurement of the systolic blood pressure was completed.
Either the connection with the hose has been interrupted or no pulse could
be determined.
Check the connections between the cuff and the main unit.
ERR 2 Non-physiological pressure has been detected.
Cause: the arm was moved during the measurement.
Repeat the measurement, keeping your arm still.
ERR 3 If the cuff takes too long to inflate, it is possible that the cuff is incorrectly
positioned or that the connection with the hose is not airtight.
Check the connections and repeat the measurement.
ERR 4 The difference between systolic and diastolic pressure is abnormally high.
Repeat the measurement under relaxed conditions, following the
instructions exactly.
If the same error occurs repeatedly with further readings, please contact
MEDISANA.
ERR 5 An average value could not be calculated due to unstable conditions while
the readings were being taken.
Repeat the measuring process, keeping your arm still and not speaking

6 Miscellaneous

GB
.
21

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

51051

Table des Matières