Miscellaneous; Care And Maintenance - Medisana HGD Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 25
GB

5 Miscellaneous

5.1
Trouble-
shooting
5.2
Care and
maintenance
5.3
Disposal
18
If the display shows one of the following error messages, this means that the
unit was unable to accurately determine blood pressure.
Error messages -
Possible causes and solutions
ERR 1 - No pulse could be detected. Make sure that the cuff is correctly fitted
and that you are holding your wrist at the level of the heart.
ERR 2 - Non-physiological pressure has been detected and has influenced the
reading. Cause: the arm was moved during the measurement. Take another
reading, keeping your arm still and not speaking.
ERR 3 - The cuff takes too long to inflate.
The cause may be that the cuff is incorrectly positioned or that the connection
with the hose is not airtight.
ERR 4 - The difference between systolic and diastolic pressure is abnormally
high. Repeat the measurement under relaxed conditions, following the
instructions exactly. If the same error occurs repeatedly, please contact
MEDISANA.
ERR 5 - An average value could not be calculated due to unstable conditions
while the readings were being taken. Repeat the measuring process, keeping
your arm still and not speaking.
• Remove the batteries before cleaning the unit.
• Never use strong detergents or hard brushes.
• Clean the unit with a soft cloth, moistened with mild soapy water. Do not
let water enter the unit. After cleaning, only use the unit when completely
dry.
• Remove the batteries from the unit if you do not wish to use it for an
extended period of time, otherwise there is a risk of battery leakage.
• Do not expose the unit to direct sunlight, and protect it from dust and
moisture.
• Only inflate the cuff when it is in position around the wrist.
• Servicing and calibration (only applies in the case of commercial use) :
The unit has been calibrated by the manufacturer for a period of two years'
use. The unit must be inspected and, if necessary, re-calibrated at least once
every two years. The inspection, which is subject to a charge, may only be
carried out by an appropriate authority or authorized service personnel as
laid down in the medical product suppliers' regulations.
Dispose of your old electrical unit in an environmentally acceptable manner
when it has reached the end of its service life!
Packaging is re-usable or may be re-cycled; metal parts must be disposed of
as scrap metal; plastics, electrical and electronic components must be disposed
of as electro-scrap.
Please remove the batteries before disposing of the equipment. Do not dis-
pose of old batteries with your household waste, but dispose of them at a
battery collection station at a recycling site or in a shop.
Consult your local council or your dealer for recycling information.

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

51422

Table des Matières