6.3
Care and
maintenance
Err 5
An average value could not be calculated due to unstable conditions while
the readings were being taken.
Repeat the measuring process, keeping your arm still and not speaking.
Err 6
There were too many errors during the measurement in MAM mode,
making it impossible to obtain a final result.
Read through the checklist for performing reliable measurements and
then repeat the measurement.
HI
The pressure in the cuff is too high (over 300 mmHg) or the pulse is too
high (over 200 beats per minute).
Relax for 5 minutes and repeat the measurement.
LO
The pulse is too low (less than 40 beats per minute).
Repeat the measurement.
Remove the batteries before cleaning the unit.
Ÿ
Never use strong detergents or hard brushes.
Ÿ
Clean the unit with a soft cloth, moistened with mild soapy water. Do
Ÿ
not let water enter the unit. After cleaning, only use the unit when
completely dry.
Remove the batteries from the unit if you do not wish to use it for an
Ÿ
extended period of time, otherwise there is a risk of battery leakage.
Do not expose the unit to direct sunlight, and protect it from dust and
Ÿ
moisture.
Only inflate the cuff when it is in position around the upper arm.
Ÿ
Servicing and calibration
Ÿ
The unit has been calibrated by the manufacturer for a period of two
years' use. In the case of commercial use, the unit must be serviced
and recalibrated at least once every two years. This calibration will
be charged for and can be carried out by an appropriate authority or
an authorised service centre - in accordance with the regulations for
users of medical equipment.
6 Miscellaneous
GB
25