Medela Invia Motion Mode D'emploi page 22

Masquer les pouces Voir aussi pour Invia Motion:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 62
Visual
What
symbol
does this
displayed
symbol
mean?
There
is a
Blinking!
blockage
+
present
acoustic
signal!
There is
a leak
present
Blinking!
+
acoustic
signal!
22
What do I have to do?
Mute the acoustic signal by pressing the
button.
Check the canister tubing and the dressing
tubing to ensure tubing is not twisted, kinked or
clamped. Verify that the correct canister is used:
087.4016. If the used canister is not the cor-
rect one contact your healthcare professional.
The notification symbol and the acoustic signal
may still be active, even if situation is resolved.
Wait for some time to let the pump react.
If you are unable to eliminate the blockage,
switch the Invia Motion NPWT pump off by
pressing the
button and holding for 3 sec-
onds. Contact your healthcare professional for
instructions.
Mute the acoustic signal by pressing the
button.
Check the dressing for leaks and pat the trans-
parent film ( clear bandage ) placed over the
wound and surrounding skin. Make sure the
Invia Motion tubing quick-connector is fully
engaged.
The notification symbol and the acoustic signal
may still be active, even if situation is resolved.
Wait for some time to let the pump react.
If you are unable to eliminate the leak, switch the
Invia Motion NPWT pump off by pressing the
button and holding for 3 seconds. Contact your
healthcare professional for instructions.

Publicité

Table des Matières
loading

Table des Matières