6.2.5. USE OF PRODUCT THROUGH THE PRESSURE OUTLET.
Ensure that the fl ow control knob is on the ZERO position (if any).
Ensure the accessory IS NOT connected to the pressure outlet.
Slowly open the cylinder valve by turning the hand wheel in anticlockwise
direction approx 1 to 1½ turns.
Sudden opening of the cylinder valve could result in a danger of fi re or
explosion arising from oxygen pressure shocks. Insu cient opening of the
cylinder valve could reduce actual fl ow delivered.
Connect the accessory to the pressure outlet.
AFTER COMPLETION OF THE THERAPY
Turn off the cylinder valve by turning the hand wheel in a clockwise direc-
tion to the stop position. Do not use excessive force.
Vent gas pressure from downstream equipment.
Disconnect the male QC probe from the pressure outlet.

6.3. AFTER USE

Turn off the cylinder valve by turning the hand wheel in a clockwise direc-
tion to the stop position. Do not use excessive force.
Reset the fl ow control knob on the "ZERO" position – when the gas venting
has ceased (valid for version with fl ow-metering device only).
Ensure the pressure indicator does not show any residual pressure.
Remove connections from user outlets.
Refi t pressure outlet and fl ow outlet protection caps. Before refi tting the
caps, ensure they are clean.
7. CLEANING
Remove dirt with a soft cloth damped in oil free soap water & rinsed with
clean water. Disinfection can be carried out with an alcohol-based solution
(with damped wipes). If other cleaning solutions are used, check that
they are not abrasive and they are compatible with the product materials
(including labels) and gas (convenient cleaning solution - i.e. Meliseptol)
Do not use cleaning solutions containing ammonia!
Do not expose to water or any other liquid.
Do not expose to high temperature (such as autoclave).
To apply the cleaning solution do not spray it as the spray may enter into
the inner parts of product and cause contamination or damage.
Do not use pressure wash as it could damage or contaminate the pro-duct.
If the inner parts of the product have been contaminated do not continue
to use the product under any circumstances. It must be withdrawn from
service.