Perform visual and audible check for possible leakages:
regulator inlet connection to cylinder valve
pressure indicator/sensor to regulator body
pressure relief valve vent hole(s)
fl owmeter (if any)
Turn off the cylinder valve by turning the hand wheel in an clockwise direc-
tion to stop position. Do not use excessive force.
If any leakage is detected, use the procedure in chapter 6.3 and return the
product to GCE for service.
6.1.4. FUNCTIONAL CHECKS BEFORE USE
Ensure the fl ow control knob is on the "ZERO" position.
Ensure the cylinder valve is open – in the "ON" position.
Check that the gauge indicates pressure/contents. If the pointer reaches
the red area send the cylinder for the fi lling
For regulators fi tted with a fl ow-metering device check that there is gas fl ow
at each setting (for instance, by listening for the sound of gas fl ow or check-
ing presence of bubbles in a humidifi er).
Turn off the shut off cylinder valve by turning the hand wheel in a clockwise
direction to the stop position. Do not use excessive force.
Reset the fl ow control knob to on the "ZERO" position once the gas fl ow
stops and the regulator is vented..
For regulators fi tted with a pressure outlet, ensure it is functional by con-
necting and disconnecting a male QC probe.
6.2. USER OUTLET S CONNECTION & USE
6.2.1. LIST OF RECOGNISED ACCESSORIES
TO BE CONNECTED TO THE FLOW OUTLET:
Humidifi ers, breathing masks or cannulas, gas savers, nebulizers.
TO BE CONNECTED TO THE PRESSURE OUTLET:
Flexible hoses, fl ow meters, Venturi suction ejectors.
At regulators fi tted with pressure outlet together with ejector outlet don´t
use quick coupler and ejector outlet in the same time. Especially when
inlet pressure is bellow 50 bar it may negatively a ect performance of
the regulator.
Before connecting any accessory or medical device to the regulator,
always check that it is fully compatible with the product connection featu-
res & the product performances.
6.2.2. PRESSURE OUTLET CONNECTION
PRESSURE OUTLET I
Ensure the male quick coupler is compatible with the pressure outlet fea-
ture.
Connect the male quick coupler.
Check if the male quick coupler is fully engaged.
7/195