Alarm Messages, continued
Severity
Alarm Header
Placement Signal Not
Reliable
(Impella 5.0 and Impella LD)
Placement Signal Not
Reliable (Impella RP)
Purge Volume Critically Low
Reinstall Software
Suction
AC Power Disconnected
Audio Off
Flight Mode Enabled
Impella Flow Reduced
Impella Position Unknown
(Impella 2.5 and Impella CP)
Impella Position Unknown
( Impella 2.5, Impella CP, Impella
5.0, and Impella LD)
Placement Monitoring
Disabled
Placement Signal Lumen
Blocked
( Impella 2.5 and Impella CP)
Preventing Retrograde Flow
Purge Cassette Incompatible
Purge Flow Decreased
Purge Flow Increased
Automated Impella Controller
Action
Placement and suction monitoring are suspended.
Flow calculation is disabled.
Refer to flow estimates in table below.
Monitor patient hemodynamics.
1. Press the Purge Menu soft key
2. then select Change Purge Fluid Bag.
Software installation was unsuccessful. Reinstall
software.
1. Reduce P-level.
2. Check Impella position.
3. Check filling and volume status.
Controller is running on battery power.
The audio for the following alarm has been
disabled.
<Alarm will be listed here>
1. Connect controller to ground during air transport.
2. If equipped with Impella Connect, enable Flight
Mode on module.
3. Upon arrival at receiving hospital, disable Flight
Mode under MENU.
1. Reduce P-Level
2. Check Impella position.
3. Check filling and volume status.
Impella position unknown due to low pulsatility,
assess cardiac function.
Confirm Impella position with imaging.
Monitor Impella position with imaging or re-enable
Placement Monitoring under MENU and Settings/
Service.
Placement and Suction Monitoring are suspended.
Aspirate with syringe, then flush.
Impella P-level has increased to prevent retrograde
flow.
1. Consider increasing target P-level.
2. For weaning, disable Retrograde Flow Control
through MENU soft key.
Contact Abiomed Service to update Impella
Controller.
The purge flow has decreased by
2.5 mL/hr or more.
This is a notification only; no action is required.
The purge flow has increased by
2.5 mL/hr or more.
This is a notification only; no action is required.
Cause
There is a problem with the Impella Catheter sensor
signal.
There is a problem with the Impella Catheter sensor
signal.
There are 15 mL (in addition to 5% of the starting bag
volume) or fewer remaining in the purge fluid bag.
Software was not installed successfully.
Suction is detected.
AC power was disconnected.
User has disabled audio for Placement Signal Not
Reliable, Purge Pressure High, Purge System Blocked,
Suction, or Placement Signal Lumen Blocked alarm.
Flight mode has been enabled for transport.
Motor speed has been reduced in response to suction.
Impella Catheter position unknown due to low
pulsatility.
Impella Catheter position unknown detected by
algorithm.
User has disabled Placement Monitoring.
Placement signal pulsatility is low and placement signal
amplitude is high (>130mmHg) for 30 consecutive
seconds.
Retrograde flow has been detected and minimum motor
speed has been increased to more than target P-level.
Incompatible purge cassette RFID version.
Purge flow has decreased by ≥2.5 mL/hr.
Purge flow has increased by ≥2.5 mL/hr.
13