Acoustical Signals - Handpiece

An acoustical signal is emitted
• every time a button is pressed,
• every time the light is turned ON,
• 1 time after 5 sec exposure time, 2 times after 10 sec,
3 times after 15 sec.
Exception: in continuous mode; an acoustical signal
is emitted every 10 seconds.
Two acoustical signals are emitted
• every time the sleep mode is terminated by pressing
the START button,
• every time the light is turned OFF.
A 2-sec error signal is emitted, if
• the handpiece over-heats,
• the battery lacks sufficient charge.
The acoustical signals from the handpiece can be turned
off (except for the 2-sec error signal). Follow these
instructions to turn them off. Put the handpiece in sleep
mode, e.g., by setting it in the charger. Take the device
from the charger; press first the TIME button, then the
START button. This changes the sleep mode of the
hand piece back to active and the status "Acoustical
signals activated" to "Acoustical signals deactivated".
The acoustical signals can be reactivated by following
the same procedure.

Troubleshooting

Error
The charge level
display on the hand-
piece glows red
steadily.
The power level
display of the hand-
piece flashes red.
The ongoing exposure
is interrupted (light off
signal is emitted) fol -
lowed by a 2-sec error
signal; the handpiece
switches to "sleep"
mode and resists
further activation.
Cause
Solution
The remaining battery charge
is adequate only for typical 5
10-sec exposure cycles.
Place the handpiece in the
charger and re-charge the
battery.
The battery lacks sufficient
charge.
Place the handpiece in the
charger and re-charge the
battery.
Error
The power level
Charging problem.
display of the hand-
The battery is defective or at
piece flashes red
the end of its useful life.
while the handpiece
is in the charger.
The handpiece has not
There is not enough charge
been used for a long
in the battery to turn on the
time and now it cannot
handpiece.
be turned on.
The light emission
The information about inter-
does not start when
mittent operation under the
the START button is
header "Technical Data -
pressed; an error
Handpiece" has not been
signal sounds for
observed.
2 sec.
The handpiece has become
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
During light emission
The information about inter-
in continuous mode,
mittent operation under the
an error signal sounds
header "Technical Data -
for 2 sec, the emission
Handpiece" has not been
is stopped, and the
observed.
handpiece changes to
The handpiece has become
sleep mode.
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
The LED showing the
The charge contact pins in
operating status in
the handpiece or the charger
the charger glows red
are wet.
steadily.
Cause
Solution
Replace the battery.
Place the handpiece in the
charger and re-charge the
battery.
Allow the handpiece to
cool for 3 minutes, then
start the next exposure by
pressing the START button.
Allow the handpiece to
cool for 3 minutes, then
start the next exposure by
pressing the START button.
Dry the charge contact
pins. Be careful not to
bend the flexible charge
contact pins in the charger.
7