Care and maintenance

Disconnect the unit from mains
supply before care and main-
tenance! The water inside the
pressure pot has to be changed
regularly (at least once a week),
because the lime particulates
harm the surface of the
pressure pot.
Polymax 1, 3 and 5
For the cleaning of the pressure
pot open the drain tap (24) at
the back of the unit and let the
water flow out. Remove dirt
thoroughly.
Attention: Do not use

General notes

Repairs on the Polymax units
may only be executed by
experts. Disconnect the unit
from mains supply at any rate.
Dreve company is only respon-
sible for the safety, reliability
and performance of the unit, if
- the unit is exclusively used
for the corresponding field of
application.
- operation is only carried out by
accordingly trained personnel.
scouring agents. Clean the
interior pot with fresh water,
a neutral washing-up liquid
and a soft scarf or sponge.
Afterwards let the water run
out. Close the drain tap (24)
and fill with fresh water.
The O-ring should be greased
regularly with silicone grease.
If the unit ventilates too slowly,
please check the function of
the silencer (25). Unscrew the
silencer if necessary, clean it in
an ultrasonic bath or replace it.
- the unit is operated in accor-
dance with these working
instructions.
- extensions, new adjustments,
changes or repairs are execut-
ed by persons authorised by
the company.
As we are currently developing
our products, we reserve the
right for technical changes.