P olymax 1, 1/120 °C, 3 and 5
Trouble shooting
Problem
Unit does not show
any function.
Unit does not show
any function.
Unit does not
heat.
Unit does not
heat.
Unit does not cre-
ate pressure.
Unit does not cre-
ate pressure.
Unit does not cre-
ate pressure.
Unit looses com-
pressed air under
the lid.
Unit looses com-
pressed air under
the lid.
Unit does not
exhaust.
Unit blows air out of
the security valve.
Addition only for the Polymax 1/120 °C
Problem
Lid cannot be
opened.
24
Cause
Mains supply is
not correct.
Main switch is off.
Temperature regu-
lator shows 0 °C.
Heating is
defective.
Unit does not have
compressed air.
Timer is not
turned on.
Magnetic valve is
defective.
Lid O-ring is
polluted.
Lid O-ring is
defective.
Absorbing duct is
polluted.
Input pressure is
too high.
Cause
Temperature in
the pressure tank
is too high.
Aid
Control the mains supply.
Turn on the main switch.
Adjust the temperature regulator.
Heating has to be exchanged by an
expert.
Control compressed air connection.
Turn on the timer.
Magnetic valve has to be exchanged
by an expert.
Clean and grease the lid O-ring.
Renew the lid O-ring.
Exchange the absorbing duct. The Polymax
5 does not have an absorbing duct.
Reduce the input pressure. The Polymax 5
has a reducing valve at the left back view.
Aid
Unit has to be cooled down.

Publicité