Electrical connection
22
Notice
There is a danger of lightning strikes if mounted out­
doors.
Use a sufficiently dimensioned overvoltage protection in
addition to the correct shielding (see Chapter 7.5, EMC
protection).
7.1

Connection in six-wire configuration

Fig. 7.1
Pin Assignment
7.2

Shortening the cable

If the transducer is connected to an amplifier in a six-wire
configuration, the transducer cable can be shortened as
required, without adversely affecting the measurement
accuracy.
A1460-4.0 HBM: public
gray
Sense lead (-)
black
Excitation voltage (-)
white
Measurement signal (+)
blue
Excitation voltage (+)
green
Sense lead (+)
red
Measurement signal (-)
Shielding/drain wire,
connected to housing
C16A