5. Instructions for Use (cont.)

5.26 Interrupting / aborting a cycle

NOTE: If the drying phase of the cycle is interrupted, the critical parts of the cycle are
complete and the message displaying the program name and "Cycle Complete. Drying
interrupted" will appear.
5.27 Additional user and setup menu functions
To access the user or setup menus, follow these steps:
The door of the HYDRIM is locked during the cycle. If the STOP
button is pressed, the cycle is aborted and cannot be re-started.
The message "Draining. Please wait" will appear. The unit will drain.
Then the message "Instruments not processed" will appear. Press
OK. Open the door by pressing
instruments. Exercise caution as instruments, trolley, baskets and
cassettes may be hot. The cycle must be repeated to ensure that
instruments are properly processed.
OR
Chamber cleaning
This cycle is used to periodically remove hardwater deposits from
the chamber walls and racks. Pour 1.0 litres of vinegar into the
chamber before starting the cycle. Touch "cleaning." A cleaning
cycle, similar to a normal wash cycle, will run. Failure to clean
chamber can promote corrosion.