5. instructions for Use
P2 Wash
FILLING
36 min
P2 Wash
Cycle Complete
P2 Wash
Instruments
not processed
OK
The cycle will start by draining the unit for 20 seconds. The bar
on the screen indicates the progression of the cycle. The stage
within the cycle is shown beneath the bar (eg. Filling, Rinsing).
The estimated time remaining (in minutes) is always displayed.
Please note that the time will increase or decrease depending
on the projected cycle completion.
When the cycle is complete, the program name and "Cycle
Complete" is displayed, and the unit beeps three times.
Press
to open the door. Use gloves to unload the
instruments. After all cycles, exercise caution as the
instruments, trolley, baskets and cassettes may be hot. Inspect
the instruments for debris prior to packaging / sterilizing.

5.5 interrupting / Aborting a Cycle

The door of the Hydrim is locked during the cycle. If the STOP
button is pressed, the cycle is aborted and cannot be re-start-
ed. The message "Draining. Please wait" will appear. The unit
will drain. Then the message "Instruments not processed" will
appear. Press OK. Open the door by pressing
if handling the instruments. Exercise caution as instruments,
trolley, baskets and cassettes may be hot. The cycle must be
repeated to ensure that instruments are properly processed.
Note that if the drying phase of the cycle is interrupted, the
critical parts of the cycle are complete and the message
displaying the program name and "Cycle Complete. Drying
interrupted" will appear.