Guidance and manufacturer's declaration—electromagnetic immunity
The Model 7711 is intended for use in the electromagnetic environment specified below. The customer or the user
of the Model 7711 should assure that it is used in such an electromagnetic environment.
Immunity
IEC
test
60601
test
level
Conducted
3 Veff
RF IEC
150 kHz
61000-4-6
bis 80
MHz
Radiated
RF IEC
3 V/m
61000-4-3
80 MHz
bis 2,5
GHz
NOTE 1—At 26 MHz and 800 MHz, the higher frequency range applies.
NOTE 2—These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection
from structures, objects, and people.
a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Model 7711 is used exceeds the
applicable RF compliance level above, the Model 7711 should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the scale.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
32
Compliance
Electromagnetic environment—
level
guidance
Portable and mobile RF communications equipment
should be used no closer to any part of the Model
005, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d =
d =
d =
10 Veff
150 kHz bis
where P is the maximum output power rating of the transmitter in watts
80MHz
(W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
10V/m
Field strengths from fixed RF transmitters, as determined by an
26MHz bis 2,7
electromagnetic site survey, should be less than the compliance level in
GHz
each frequency range. Interference may occur in the vicinity of
equipment marked with the following symbol:
(
0,4 √P
0,4 √P
für 80 MHz bis 800 MHz
0,7 √P
für 800 MHz bis 2,5 GHz
)
)
(
)
(