5.5 Taking Voltage Readings

Read Feeding Voltage
Take battery voltage feeding
PCB.
Expected result: 23 to 29 V
Take voltage feeding main
switch (two middle pins)
Expected result: 23 to 29 V
Take voltage feeding the lift
motor whilst pressing the UP
or DOWN button on the hand
control.
Expected result: 23 to 29 V
1586747-A
Read Feeding Voltage
Take voltage feeding the hoist
mover PCB. (Robin Mover™
only)
Expected result: 23 to 29 V
Take voltage feeding the copper
rings whilst pressing the LEFT
or RIGHT button on the hand
control. (Robin Mover™ only)
Expected result: 23 to 29 V
Take voltage feeding the trolley
mover PCB whilst pressing the
LEFT or RIGHT button on the
hand control. (Robin Mover™
only)
Expected result: 23 to 29 V