Switch TMP Alarm (Low) on
1. Turn on the unit and press the [>] button for about 3 seconds until the LEDs turn green.
2. Press the [<] or [>] button to set the alarm value
3. Press [HOLD] to save the setting or [Power] to return to normal measurement mode without saving.
9. Calibration mode
Use a reference oil whose TPM is known for calibration
1. Heat the calibration oil to 50 ° C
2. After turning on the calibration mode (see Config mode), CAL is displayed. Insert the sensor into the calibration oil,
paying attention to the Min / Max markings on the sensor shaft.
3. If the reading is stable, press the [<] and [>] buttons to adjust the reading in 0.5% increments.
4. If the display value matches the reference value after setting, press the [HOLD] key to store the value.
10. Call up the factory settings
1. Use the configuration menu to enter the factory settings (RST).
2. Press the [>] and [HOLD] buttons at the same time to reset your calibration to factory defaults.
11. Statutory Notification about the Battery Regulations
The delivery of many devices includes batteries, which for example serve to operate the remote control. There also
could be batteries or accumulators built into the device itself. In connection with the sale of these batteries or
accumulators, we are obliged under the Battery Regulations to notify our customers of the following:
You can obtain further information about the Battery Regulations from the Bundesministerium für Umwelt, Naturschutz
und Reaktorsicherheit (Federal Ministry of Environment, Nature Conservation and Reactor Safety).
All rights, also for translation, reprinting and copy of this manual or parts are reserved. Reproduction of all kinds
(photocopy, microfilm or other) only by written permission of the publisher.
This manual considers the latest technical knowing. Technical alterations reserved.
PeakTech provides a three-year warranty on all devices from the date of purchase.
We hereby confirm that all devices, the specifications mentioned in our documentation, are delivered and calibrated at
the factory.
An annual repetition of the calibration over a calibration laboratory is recommended.
Please dispose of old batteries at a council collection point or return them to a local shop at
no cost. The disposal in domestic refuse is strictly forbidden according to the Battery
Regulations. You can return used batteries obtained from us at no charge at the address on
the last side in this manual or by posting with sufficient stamps.
Batteries, which contain harmful substances, are marked with the symbol of a crossed-out
waste bin, similar to the illustration shown left. Under the waste bin symbol is the chemical
symbol for the harmful substance, e.g. „Cd" for cadmium, „Pb" stands for lead and „Hg" for
mercury.
13