Technical Specifications - Dräger Pac 3500 Notice D'utilisation

Masquer les pouces Voir aussi pour Pac 3500:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 16
War-
Cause
ning
code
160
Date and time are adjusted after
software update or battery
replacement
162
Set operating time of the instru-
ment almost expired
222
Calibration interval expired
242
Bump test interval expired
10 Changing dust and water filter
Remedies
Set date and time with Dräger
CC-Vision
Set new operating time of the
instrument
Calibrate instrument
Perform bump test or calibration

11 Technical specifications

Environmental Conditions
o
During operation
–30 to 50
C / –22 to 122
700 to 1300 hPa
10 to 90% relative humidity
o
Conditions for stor-
0 to 40
C / 32 to 104
age
30 to 80% relative humidity
Ingress protection
IP 68
Pac 3500
2 years (typical at 25
Operating times
Pac 5500
unlimited
Operating times
Battery life (typical at
8 hours of use per day,
o
25
C)
1 minute alarm per day:
CO, H
S: > 2 years typical
2
O
: > 12 month typical
2
Intensity of alarm
typical 90 dB (A) at a distance of 30 cm (11.8 in. )
Dimensions
64 x 84 x 20 (battery compartment 25) mm
2.5 x 3.3 x 0.8 (battery compartment 1) in.
Weight
106 g (3.8 oz.)
Approvals
(see "Notes on Approval" on page 195)
o
F
o
F
o
C)
7

Publicité

Table des Matières
loading