Storage Conditions; Ec Declaration Of Conformity - Mettler Toledo InPro 7100 Instructions D'utilisation

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 18
en
14
Only use cleaning agents that are compatible with
the materials to be cleaned. In addition, do not use
coarse materials such as steel wool for cleaning,
as they will damage the surface.
Rinse the conductivity cell thoroughly with distilled
water before installation.
7
Troubleshooting
1. If no signal appears at the display unit, check
the electrical connections. Another possible
cause of this problem is that the conductivity
cell is not in contact with the medium.
2. If no temperature signal is present, the tempera-
ture sensor may not be connected properly or
the device may not support NTC temperature
sensor (for InPro 7100 i) or Pt1000 temperature
sensor (for InPro 7100).
3. If the measuring signal is incorrect or instable,
the cause may be that the conductivity cell is
contaminated or fitted too close to the pipe (the
distance between the pipe and the end of the
conductivity cell should be 30 mm).
8
Disposal
The user is responsible for the proper disposal of
the conductivity sensor. The sensor contains elec-
tronic components that must be disposed of cor-
rectly in order to avoid hazards to people or to the
environment.
9
Warranty
Warranted free of manufacturing defects for a term
of 12 months after delivery.

10 Storage conditions

If the conductivity cell is not in use, it can be stored
dry at room temperature.

11 EC declaration of conformity

The complete declaration is available in the dec-
laration of conformity certificate.
InPro and ISM are trademarks of the Mettler-Toledo
Group.
© 02 / 16 Mettler-Toledo GmbH
Printed in USA