2.2.5 3D Kinetics
With the 3D kinetics tab, conduct a kinetics follow-up for all wavelengths depending on
time.
The tab contains several areas:
By configuring times between two acquisitions, the software gives you the maximum
acquisition time. In fact, the device is able to acquire up to 50 curves.
Each curve has an instant t.
If you so wish, you can enter a deadline before starting the acquisition.
If you don't know the reaction time to be observed, you can work in continuous acquisition.
In which case, two modes are suggested:
-
The device acquires 50 curves, and then, without any action from the experimenter, it
continues the acquisitions by deleting a curve for each new moment acquired.
-
The device acquires 50 curves, and then, without any action from the experimenter, it
continues the acquisitions by deleting a curve for each new moment acquired.
Warning:
be multiplied by 2 for each acquisition cycle.
The graphical area contains 4 areas.
ENGLISH
-
Step 1: Calibration (similar to previous tabs)
-
Step 2: Adjusting the acquisition parameters
By deletion of the first curves.
By deletion of one curve out of two
This implies that the time between two acquisitions will be modified. In fact, it will
24