Guidance and manufacturer's declaration—electromagnetic immunity
The S8 series of devices is intended for use in the electromagnetic environment specified below. The
customer or the user of the S8 device should assure that the device is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines.
IEC 61000-4-11
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE: Ut is the AC mains voltage prior to application of the test level.
42
All manuals and user guides at all-guides.com
IEC60601-1-2
test level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/
output lines
±1 kV differential
mode
±2 kV common
mode
<5% Ut
(>95% dip in Ut)
for 0.5 cycle
40% Ut
(60% dip in Ut)
for 5 cycles
70% Ut
(30% dip in Ut)
for 25 cycles
<5% Ut
(>95%dip in Ut)
for 5 sec
3 A/m
Electromagnetic
Compliance level
environment—guidance
±6 kV contact
Floors should be wood,
concrete or ceramic tile. If
±8 kV air
floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
±2 kV
Mains power quality should
be that of a typical
commercial or hospital
Not Applicable
environment.
±1 kV differential
Mains power quality should
mode
be that of a typical
commercial or hospital
±2 kV common
environment.
mode
< 12 V
Mains power quality should
(>95% dip in 240V)
be that of a typical
for 0.5 cycle
commercial or hospital
environment.
96 V
If the user of the S8 device
(60% dip in 240V)
requires continued operation
for 5 cycles
during power mains
interruptions, it is
168 V
recommended that the S8
(30% dip in 240V)
device be powered from an
for 25 cycles
uninterruptible power
source.
<12 V
(>95%dip in 240V)
for 5 sec
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.