7.4 Timing functions
7.4.1
Timer
The timer function stops the distillation after a set period.
A set timer is displayed by inserting an arrow "down" below the
min. display.
If a timer is not set, the timer is always counted upward. The
arrow "up" is inserted.
1.
Activate the timing display with the aid of the selection window by using the rotary
pushbutton controller Hei-GUIDE (9).
2.
The selection window flashes when pressing the rotary pushbutton Hei-GUIDE (9).
The value can therefore be adjusted. A time between 1 and 999 minutes can be se-
lected by turning the Hei-GUIDE (9).
3.
Turn the rotary pushbutton controller Hei-GUIDE (9) to set the display value to the
desired timing value.
4.
Pressing the Hei-GUIDE (9) again takes over the selected new value. The evapora-
tor responds immediately and the time is counted backward.
5.
An arrow "down" (3) will be inserted under the minute display.
The time is running backwards at the start of the rotation with the Rotation start (8)
pushbutton. The distillation is interrupted after completing the set time.
−
the timer display shows "000" with a flashing frame
•
Motor lift version:
−
The rotation is switched off.
−
The lift is raised (only Hei-VAP Advantage, Hei-VAP Precision).
−
The heating bath is switched off.
•
Hand lift version
−
The rotation changes to 20 rpm till manual intervention by the user.
−
The heating bath is switched off.
The screen is reset to value "---" when activating the Hei-GUIDE
(turning or pressing), the frame flashes, and the value can be
change or confirmed.
If value "---" is confirmed by pressing the Hei-GUIDE, the timer
will count the run time upward after the start of the rotation. The
up arrow will be illuminated. This value and the frame will flash
when reaching the maximum value of "999". The process will
continue.
The value is set at zero and the "timer" will again count upward
when activating the Hei-GUIDE (turning or pressing), if the
process has not been stopped.
7.5 Error messages
Error messages are only displayed in the event of an error. The error codes 1 to 6 and
their troubleshooting are explained in chapter 8, "Errors and Troubleshooting".
7.6 Upgrading
An upgrade from the evaporator model Hei-VAP Advantage to
evaporator model Hei-VAP Precision is possible by changing the
control panel and adding a vacuum box.
Hei-VAP Value, Hei-VAP Advantage
Operation of the Hei-VAP Advantage
Timer
EN
51