WARNINGS
‣ Do not bring the device or accessories into a Magnetic Resonance (MR) environment as it may cause
unacceptable risk to the patient or damage to the oxygen concentrator or MR medical devices. The device and
accessories have not been evaluated for safety in an MR environment.
‣ Do not use the device or accessories in an environment with electromagnetic equipment such as CT scanners,
Diathermy, RFID and electromagnetic security systems (metal detectors) as it may cause unacceptable risk to the
patient or damage to the oxygen concentrator. Some electromagnetic sources may not be apparent, if you notice
any unexplained changes in the performance of this device, if it is making unusual or harsh sounds, disconnect
the power cord and discontinue use. Contact your healthcare provider or clinical staff.
‣ This device is suitable for use in home and healthcare environments except for near active HF surgical equipment
and the RF shielded room of a magnetic resonance imaging system, where the intensity of electromagnetic
disturbances is high.
‣ Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in
improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
‣ Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the oxygen concentrator, including cables specified
by the manufacturer. Otherwise, degradation of the performance of this equipment could result.
‣ Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
EN - 22
SE-1060AW