SYMPTOM
F. Unit operates. Red Low Flow LED is
illuminated. Audible alert may be sounding.
G. Red Service Required LED is illuminated
and an intermittent audible signal is
sounding.
H. Unit operates. Audible low-frequency
vibration sound is detected.
I. Unit operates. Any of the visual and
audible alerts do not function when the
device is turned ON
J. Red Service Required LED flashes, no
AC power present, with audible alarm
K. If any other problems occur with your
oxygen concentrator.
5.0 Specifications
Catalog Number
Delivery Rate
Maximum Recommended Flow
(@ nominal outlet pressures of zero
and 7 kPa)**
Outlet Pressure
Oxygen Percentage
Minimum volume fraction of
delivered oxygen at the maximum
rated flowrate
Flow Meter Accuracy
5.1 ENVIROMENTAL
Operating Atmospheric Pressure
Operating Temperature Range
Operating Relative Humidity Range
Power Consumption
Sound Pressure Level at 3 and 10
LPM
Sound Power Level at 3 and 10
LPM
5.2 PHYSICAL
Weight
Dimensions
Transport and Storage Conditions
SE-1060AW
POSSIBLE CAUSE
1. Blocked or defective
cannula, face mask, or oxygen
tubing.
2. Blocked or defective
humidifier bottle.
3. Flow meter set too low.
1. Electronic assembly or other
internal component
malfunction.
1. Electronic assembly
malfunction.
1. Electronic assembly
malfunction.
1. Power Failure Alarm
PulmO2 10 LITER OXYGEN CONCENTRATOR
-25˚C (-13˚F) to 70˚C (158˚F), humidity range of 15% to 95% non-condensing
1. Detach cannula or face mask. If proper flow is restored,
clean or replace if necessary. Disconnect the oxygen tubing at
the oxygen outlet. If proper flow is restored, check oxygen
tubing for obstructions or kinks. Replace if necessary.
2. Detach the humidifier from the oxygen outlet. If proper flow
is obtained, clean or replace humidifier.
3. Set flow meter to prescribed flow rate.
If the above remedies do not work, contact your service
technician.
1. Turn your unit OFF. Call your service technician and switch
to alternate oxygen source
1. Turn your unit OFF. Call your service technician and switch
to alternate oxygen source.
1. Turn your unit OFF. Call your service technician and switch
to alternate oxygen source.
1. Turn unit off until electricity or power is restored.
1. Turn your unit OFF. Call your service technician and switch
to alternate oxygen source.
1060AW
1 to 10 LPM
10 LPM
58.6 ± 3.5 kPa (8.5 ± 1.0 PSIG)
1-10 LPM = 93%-6%, +3%
85% O2 @ 10 LPM
+/- 0.5 SLPM, at 10 SLPM; +0, -0.5 SLPM
1013 hPa to 795 hPa
0-2000 meters (0-6562 feet)
Tested at nominal voltage.
5˚C (41˚F) to 40˚C (104˚F)
15% to 95%, non-condensing
230 VAC, 50Hz: 520 Watts Maximum
<60 dBA
<70 dBA
25.2 kg (55.5 lbs)
70.5 cm H x 40.6 cm W x 35.6 cm D
(27.75" H x 16" W x 14" D)
REMEDY
EN - 19