Catalog Number
5.3 STANDARDS COMPLIANCE
Approval Body and Safety Standard
EMC Compliance To
5.4 ELECTRICAL
Electrical Rating
5.5 PRESSURE
Maximum Limited Pressure
under Normal condition
under Single Fault condition
Operating System
5.6 CLASSIFICATION
Equipment Class and Type
The equipment power consumption is greatest within the first 150 hours of operation, and will decrease after this initial break-in
period.
Specifications subject to change without notice.
Units of measure for volumes, flows and leakages are expressed in STPD: 101,3 kPa at an operating temperature of 20 °C, dry
Oxygen Concentration by Flow Rate
Flow
%O
L/m
2
10
93% -6%, +3%
9
93% -6%, +3%
8
93% -6%, +3%
7
93% -6%, +3%
6
93% -6%, +3%
5
93% -6%, +3%
4
93% -6%, +3%
3
93% -6%, +3%
2
93% -6%, +3%
1
93% -6%, +3%
5.7 Service Life
The expected service life of the PulmO2 10 L oxygen concentrator, which includes the performance of any required service is 5
years. Drive DeVilbiss recommends for optimal service life that the PulmO2 10L Oxygen Concentrator be operated for at least 30
minutes after it is powered on. Shorter periods of operation, operating in extreme temperature/humidity conditions or in the
presence of contaminates, and/or handling and storage conditions outside those specified, may affect the long term reliable
EN - 20
PulmO2 10 LITER OXYGEN CONCENTRATOR
1060AW
TUV
IEC 60601-1:2020
IEC 60601-1-6:2020
IEC 60601-1-11:2020
EN ISO 80601-2-69:2020
EN60601-1-2:2020
100-240 VAC, 50 / 60Hz, 3.2 Amp
186 kPa (27 PSIG)
276 kPa (40 PSIG)
Time Cycle / Pressure Swing
Class II Equipment;
Type BF Applied Part
IP21 Ingress Protection
SE-1060AW