Product Components; Installation - Hach 2200 PCX Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Product components

Make sure that all components have been received. Refer to
analog input/output kit, water weir and software, refer to the expanded manual on the manufacturer's
website.. If any items are missing or damaged, contact the manufacturer or a sales representative
immediately.
Figure 2 Product components
1 Particle counter with power cord
2 Cleaning solution
3 Cleaning brush

Installation

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Installation guidelines
Install the instrument:
• As near the sample source as possible to decrease analysis delay
• In a clean, dry, well ventilated and temperature controlled location
• In a location with minimum vibrations that has no direct exposure to sunlight
• In an environmental enclosure that supplies protection from precipitation and direct sunlight, good
ventilation and temperature control if installed outdoors
• In a location where the power switch and power cord are visible and easily accessible
• In a location where there is sufficient clearance around it to make plumbing and electrical
connections
Figure
4 Power cord
5 Power supply
W A R N I N G
2For optional items such as the
English 7